  data sheet zrh this ndir gas analyzer is used for measuring co 2 , co, so 2 and ch 4 and features a high accuracy, multiple functions and easy operation through use of a microprocessor. it is housed in an easy-to-use horizontal case (can be flush mounted on panel, mounted on 19 inch rack, placed on table top), or vertical case (can be flush mounted on panel, mounted on wall). the analyzer provides a performance superior to the con- ventional double-beam system, is easy to maintain, and offers an excellent long-term stability. it is thus optimum for con- tinuous measurement in the combustion control of various industrial furnaces, in research on garden-plants, and so on. 1. the use of a microprocessor provides high accuracy, multiple functions and easy operation. zero and span calibration is accurate and easy just by pressing the calibrating keys. a self-diagnosis function is included. an automatic calibrating function can be provided as an option. range can be changed over by a external signal as an option. 2. an improved optical system provides long term stabil- ity, and there is a minimum of drift caused by contami- nation of the measuring cell, so the long-term stability is excellent. 3. adopting a serial dual-layer type of transmission detec- tor minimizes remarkable the interface from other gas components. 4. easy maintenance. the single-beam photometric system uses a sample cell only and eliminates the necessity of delicate adjustment for optical balance. the instrument is designed as a unit of simple construction featuring easy maintenance and checks. measurable gas components: single-component co 2 co, so 2 , ch 4 dual-component co 2 /co: landfill garbage application co 2 : 0 to 50% ch 4 : 0 to 80% measuring range: refer to measurable range combination table (page 5). measuring system: non-dispersion infrared-ray absorption method, deviation method, single light source ?single beam output signal: output 1; 0 to 1v dc output 2; 4 to 20ma dc (optional allow- able load resistance 550 ? or less). repeatability: 1st range (low range) within 0.5% of full scale 2nd range (high range) within 1% of full scale linearity: 2% of full scale zero drift: within 2% of full scale/week span drift: within 2% of full scale/week response time: within 15 seconds max. (for 90% re- sponse) depending on cell length power supply: 100v, 115v, 220v or 240v 10% ac, 50/ 60hz power consumption: 37va max. ndir type infrared gas analyzer features specifications date EDS3-111E nov. 24, 2004
zrh 2 ambient temperature: ? to +45 c ambient humidity: 90% rh or less enclosure: steel casing, for indoor use storage condition: temperature; ?0 to +60 c humidity; 100% rh max. (free from con- densation) outer dimensions (h x w x d): (1) horizontal case rack mounting type; 133 x 483 x 435 mm panel flush mounting type; 133 x 443 x 435 mm table-top type; 145 x 443 x 435 mm (2) vertical case panel flush mounting type; 440 x 294 x 178 mm wall mounting type; 484 x 294 x 178 mm mass weight: approx. 12kg finish color: munsell 2.5y8.4/1.2 display: 4 digit led for concentration display 4 digit led for sub-display output hold: output value before manual or automatic calibration is hold. whether or not to ef- fect hold function can be selected. sample gas condition: temperature; 0 to 50 c dust; less than 0.3 m pressure; less than 10 kpa standard adjustment gas: dry n 2 balance warm up time: approx. 2 hours material of gas-contacting parts: sample cell, sus304, neoprene rubber infrared-ray transmitting window; caf 2 or sapphire internal tubing; toaron tube gas inlet/outlet, purge gas inlet size: rc1/4 (pt1/4 internal thread) or npt1/4 internal thread measured gas flow rate: 1 0.5 liter/minutes purge gas flow rate: approx. 1 liter/minute scope of delivery :analyzer, power fuse, manual, mounting bracket in the case of panel mounting type mounting method: (1) horizontal case mounted on 19 inch rack, or on panel, or on table-top remark: 70% or more of the analyzer weight should be supported at the bottom of the case. (in case of mounting panel or 19 inch rack, provide a support at the rear of casing.) installation conditions: install the analyzer at a place not exposed to direct sunlight or the radiation from a high temperature object. a void vibra- tion, and select a clean place free of cor- rosive and/or combustible gases. if in- stalling outdoors, provide a suitable cas- ing or cover to protect the analyzer from wind, rain, etc. approx. 250mm 90 support optional specifications remote output hold: analog output (dc0-1v, 4-20ma) is held via external signal. input signal: 5v dc remote range changeover: range is changeable via external signal. range changeover input signal: 5v dc range identification signal output: contact output; 1 a contact contact capacity; 250v ac, 2a (resistive load) automatic calibration: zero and span are automatically calibrated at the preset cycle. calibration gas is supplied sequentially by driving an elec- tromagnetic value installed outside. calibration channel: up to 2 components can be calibrated si- multaneously. zero calibration point: fixed at 0% span calibration point: 50 to 100% full scale calibration start: via built-in timer or remote start signal output hold at calibration: possible calibration gas flow mode: (1) zero gas (2) zero gas ?span gas 1 (3) zero gas ?span gas 2 (4) zero gas ?span gas 1 ?span gas 2 calibration gas flow time: settable from 100 to 599 sec. calibration cycle: 1 to 199 hours (in 1-hour step) calibration failure alarm: provided when fault occurs during auto calibration. contact output: during calibration; 1 a (n.o) contact, con- tact capacity 250v ac, 2a (resistive load) calibration failure; 1 a (n.o) contact, con- tact capacity 250v ac, 2a (resistive load) electromagnetic valve drive; 1 a (n.o) con- tact, contact capacity 250v ac, 2a (re- sistive load) remote start: remote start signal; voltage input 5v dc (2) vertical case flush mounted on panel or mounted on wall 90 90 or
3 notes : (1) refer to measurable range combined table (2) output signal are provided simultaneously (3) 0 to 100mv, 0 to 10mv dc is available on request (4) to be adviced components of sample gas (1) single-component analyzer code symbols -- 2 yy h r za- 14 15 y 16 17 18 9 10111213 2 1 345678 description no. of measurable components single component horizontal case single component vertical case measuring range (2nd range) (note 1) none 0 to 1000 ppm 0 to 2000 ppm 0 to 2500 ppm 0 to 5000 ppm 0 to 1% 0 to 2% 0 to 5% 0 to 10% 0 to 20% 0 to 50% 0 to 100% others paint fuji standard structure table top type (for horizontal case only) 19 inch rack mounting type ( for horizontal case only ) panel mounting type wall mounting type measuring range (1st range) (note 1) 0 to 500 ppm 0 to 1000 ppm 0 to 2000 ppm 0 to 2500 ppm 0 to 5000 ppm 0 to 1% 0 to 2% 0 to 3% 0 to 5% 0 to 10% 0 to 20% 0 to 40% 0 to 50% 0 to 70% 0 to 100% others power supply, piping connection 100v ac 50hz rc 1/4 100v ac 60hz rc 1/4 115v ac 60hz rc 1/4 220v ac 50hz rc 1/4 100v ac 50hz npt 1/4 100v ac 60hz npt 1/4 115v ac 60hz npt 1/4 220v ac 50hz npt 1/4 240v ac 50hz npt 1/4 240v ac 50hz rc 1/4 adjustment gas standard for carburizing furnace others (note 4) non standard specifications non standard output signal (output 1/2) (notes 2, 3) 0 to 1v dc /4 to 20ma dc 1 3 e f g u h j k q l m n w p x r z a b c d y a y a 0 1 2 3 4 5 6 7 8 9 a y f g u h j k l m n p r z a b z y z a b d e measurable component so 2 co co 2 ch 4 optional function (2) none with remote range, range identification and remote output hold optional function (1) none with auto-calibration
zrh 4 the product conforms to the requirements of the electromagnetic compatibility directive 89/336/eec as detailed within the technical construction file number tz734576. the applicable standards used to demonstrate compliance are : en 50081-1 : 1991 class a conducted and radiated emissions en 50082-1 : 1992 radiated immunity, esd and fbt (2) dual-component analyzer note (5) refer to measurable range combination table (6) 1st comp. is co 2 , 2nd comp. is co. note: available for zrh1, 2 -- 2 h r zg a- 14 15 16 17 18 9 10111213 2 1 345678 description no. of measurable components dual component horizontal case dual component vertical case structure table top type 19 inch rack mounting type panel mounting type wall mounting type measuring range (1st comp. 1st range) note (5, 6) 0 to 5000 ppm 0 to 1% 0 to 2% 0 to 5% 0 to 10% 0 to 20% 0 to 50% 0 to 100% others measuring range (2nd comp. 1st range) note (5, 6) 0 to 500 ppm 0 to 1000 ppm 0 to 2000 ppm 0 to 5000 ppm 0 to 1% 0 to 2% 0 to 5% 0 to 10% 0 to 20% 0 to 25% 0 to 50% 0 to 100% others output signal (output 1/2) (notes 2, 3) 0 to 1v/dc 4 to 20ma dc 2 4 h j k l m n p r z a b c d y a y a e f g h j k l m n v p r z power supply, piping connection 100v ac 50hz rc 1/4 100v ac 60hz rc 1/4 115v ac 60hz rc 1/4 220v ac 50hz rc 1/4 100v ac 50hz npt 1/4 100v ac 60hz npt 1/4 115v ac 60hz npt 1/4 220v ac 50hz npt 1/4 240v ac 50hz npt 1/4 240v ac 50hz rc 1/4 0 1 2 3 4 5 6 7 8 9 a g measurable component co 2 /co optional function (2) none with remote range, range identification and remote output hold optional function (1) none with auto-calibration -- 2 h r zg a- 14 15 16 17 18 9 10111213 2 1 345678 description measuring range (2nd comp. 2nd range) none 0 to 1000 ppm 0 to 2000 ppm 0 to 5000 ppm 0 to 1% 0 to 2% 0 to 5% 0 to 10% 0 to 20% 0 to 25% 0 to 50% 0 to 100% others adjustment gas standard for carburizing furnace others y f g h j k l m n v p r z y z measuring range (1st comp. 2nd range) none 0 to 1% 0 to 2% 0 to 5% 0 to 10% 0 to 20% 0 to 50% 0 to 100% others y j k l m n p r z a b z non standard specifi- cations non standard paint fuji standard
5 e f g u h j k q l m n w p x r measurable range combination table (1) single-component (co 2 ?co?ch 4 ) 0 to 500 ppm 0 to 1000 ppm 0 to 2000 ppm 0 to 2500 ppm 0 to 5000 ppm 0 to 1% 0 to 2% 0 to 3% 0 to 5% 0 to 10% 0 to 20% 0 to 40% 0 to 50% 0 to 70% 0 to 100% f 0 to 1000 ppm g 0 to 2000 ppm u 0 to 2500 ppm h 0 to 5000 ppm j 0 to 1% k 0 to 2% l 0 to 5% m 0 to 10% n 0 to 20% p 0 to 50% r 0 to 100% :co 2 :co : ch 4 :so 2 : impossible * also single range is possible (2) dual-components (co 2 /co) h j k l m n p r 0 to 5000 ppm 0 to 1% 0 to 2% 0 to 5% 0 to 10% 0 to 20% 0 to 50% 0 to 100% co co 2 1st component 1st range 2nd component 1st range 1st range 2nd range infrared?ay light source gas inlet gas outlet detector mass flow sensor measuring cell rotary sector m synchronuous motor ac amplifier rectifier dc amplifier a/d cpu d/a operating panel contact output display operation key output 0 to 1v, 4 to 20ma dc input range identification signal output during calibration remote range remote output hold fundamental principle diagram 1st component is co 2 , 2nd component is co. ? : dual-components are possible both components are able to have 2nd range. 2nd range is x 2 or x 2.5 of 1st range, choose the code symbols * ? : dual-components are possible but only one component, co 2 or co, is able to have 2nd range. 2nd range is x 2 or x 2.5 of 1st range, choose the code symbols. : impossible. r 0 to 100% ? ? ? ? ? ? ? ? p 0 to 50% ? ? * ? ? ? ? ? ? v 0 to 25% ? ? ? ? ? ? ? ? n 0 to 20% ? ? ? ? ? ? ? ? m 0 to 10% ? ? ? ? ? ? ? ? l 0 to 5% ? ? ? ? ? ? ? ? k 0 to 2% ? ? ? ? ? ? ? ? j 0 to 1% ? ? ? ? ? ? ? ? h 0 to 5000ppm ? ? ? ? ? ? ? ? g 0 to 2000ppm ? ? ? ? ? ? ? ? f 0 to 1000ppm ? ? ? ? ? ? ? ? e 0 to 500ppm ? ? ? ? ? ? ?
zrh 6 483 465 443 133 57.2 7 7 power switch operating panel (rack mounting plate) 380 124.5 30 25 external connection terminal (1st component) (2nd component) (auto cal) purge gas inlet (rc 1/4 or npt 1/4) power terminals 435 zrh panel cutout dimensions (in case of panel mounting) 126 +2 0 436 +2 0 sample gas outlet (rc 1/4 or npt 1/4) sample gas inlet (rc 1/4 or npt 1/4) outline diagram (unit:mm) (1) horizontal case
7 (2) vertical case panel flush mounting wall mounting 178 440 424 170 148 294 118 35 35 35 30 80 280 door open -200 283 radius rc1/4 or npt1/4 rc1/4 or npt1/4 rc1/4 or npt1/4 (sample gas outlet) (sample gas inlet) (purge gas inlet) panel mounting fixture wiring hole ( 22) with rubber bushing 150 350 281 +2 0 430 +2 0 430 +2 0 panel cutout and spacing 178 484 440 424 454 170 148 294 118 10 70 35 35 35 30 80 280 5 wiring hole ( 22) with rubber bushing door open -200 283 radius rc1/4 or npt1/4 rc1/4 or npt1/4 rc1/4 or npt1/4 (sample gas outlet) (sample gas inlet) (purge gas inlet)
zrh 1 x gas analyzer main unit 1 x test report 1 x instruction manual 2 x power fuse 4 x panel mounting bracket gas sampling device accommodating locker standard gas (for calibration) receiving instrument 1. analyzer type. 2. maximum, normal and minimum concentrations of sample gas as well as type and content (percent by volume) of con- comitant gas. 3. temperatures (maximum, normal and minimum), pressure and humidity of sample gas. 4. dust conditions (mg/nm 3 or particle size, characteristics, etc.) and environmental conditions. 5. other items 17 28 39 410 511 612 17 28 39 410 511 612 17 28 39 410 511 612 output 1: 0 to 1v dc output 2: 4 to 20ma dc fault range identification contact output (option) remote hold input(option) remote range input(option) output 1: 0 to 1v dc output 2: 4 to 20ma dc range identification contact output (option) remote range input(option) under calibration contact output auto calibration failure contact output remote start input zero gas contact output span gas 1 contact output span gas 2 contact output 1st component 2nd component auto cal (option) connection diagram (same for horizontal and vertical cases) scope of delivery ordering information related devices m4 screws used information in this catalog is subject to change without notice. printed in japan head office 6-17, sanbancho, chiyoda-ku, tokyo 102-0075, japan http://www.fesys.co.jp/eng sales div. international sales dept. no.1, fuji-machi, hino-city, tokyo, 191-8502 japan phone: 81-42-585-6201, 6202 fax: 81-42-585-6187 http: //www.fic-net.jp/eng caution on safety *before using this product, be sure to read its instruction manual in advance.


▲Up To Search▲   

 
Price & Availability of EDS3-111E

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X